Medical Technologies National Congress (TIPTEKNO), Trabzon, Türkiye, 12 - 14 Ekim 2017
Electromagnetic field and treatment systems are one of the important research subjects of today. In this project, antenna and coil designs and analyzes were performed for different types of DMAT treatment systems. Electromagnetic field exposure, besides power density and distance; It is determined by physical properties such as the duration of exposure, type (near or far field), frequency of the source, magnitude of the electric and magnetic fields and modulation (continuous wave or pulse modulation). Maximum RF energy interaction in biological systems can be determined with penetration and distribution in biological tissues, energy absorption, power storage, generated electrical and magnetic fields. Their quantities vary in relation to the size and physical form of the target. The aim of this project to develop a device that provides increase of 3 te 1 in wound healing with especially in the 27.1 MHz ISM band with 10-40 duty cycle and pulse modulated at 1 kHz, with adaptive band feature. The primary goal in this study is to produce a device that accelerates wound healing.